Page 1 - Combustion Gas Analyzer

Fyrite® ProCombustion Gas AnalyzerInstruction 0024-9395Operation & MaintenanceRev. 12 – October 2010Product Leadership • Training • Service • Reli

Page 2 - WARRANTY

Instruction 0024-93954Fyrite ProMemory (Model 105/110/120/125) ... Up to 10 sets of readings can be saved in memory. When memory is

Page 3 - Contents

Instruction 0024-93955Fyrite Pro3.0 SETTING UP THE ANALYZER3.1 Preliminary StepsBefore using the analyzer . . .• Check batteries (Section 3.2)• Conne

Page 4

Instruction 0024-93956Fyrite Pro3.3 Connecting the Probe Hose and ThermocoupleThe analyzer’s model number determines which hose and thermocouple conn

Page 5 - Ordering Information

Instruction 0024-93957Fyrite ProFigure 3-2A. Connecting the 0019-7110 Probe (Model 100)GAS INLETSAMPLE-GASHOSEWATER TRAP /FILTER ASSEMBLYAC ADAPTERJAC

Page 6

Instruction 0024-93958Fyrite ProFigure 3-2B. Connecting the 0024-3004 Probe (Model 105)DIFFERENTIALPRESSURE HOSE(Optional)WATER TRAP / FILTERASSEMBLYP

Page 7 - 1.1 The Fyrite Pro

Instruction 0024-93959Fyrite ProFigure 3-2C. Connecting the 0019-7111 Probe (Model 110/120)T-STACKGASINLETSAMPLE-GASHOSETHERMOCOUPLECONNECTOR(T-STACK)

Page 8 - 1.2 Operational Overview

Instruction 0024-939510Fyrite ProFigure 3-2D. Connecting the 0024-3004 Probe (Model 125)T-STACKT-AIRROOM AIR /PRIMARY AIRTHERMOCOUPLECONNECTOR(T-AIR)D

Page 9 - 2.0 TECHNICAL CHARACTERISTICS

Instruction 0024-939511Fyrite Pro3.4 Front Panel Push ButtonsNote that a push button may perform several functions, depending on the analyzer’s model

Page 10 - Fyrite Pro

Instruction 0024-939512Fyrite Pro3.5 Setup ModeThe analyzer is preset at the factory for the parameters shown below, but can be changed as described

Page 11 - 3.1 Preliminary Steps

Instruction 0024-939513Fyrite Pro3.5.3 CO Channel SetupThe CO Channel Setup Screen is labeled “CO.” 1. Enter the Setup Mode per Section 3.5.1. If nec

Page 12 - Thermocouple

WARRANTYBacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will c

Page 13

Instruction 0024-939514Fyrite Pro3.5.5 Time SetupThere are two Time Setup Screens, one for hours and the other for min-utes. Two bars appear above th

Page 14

Instruction 0024-939515Fyrite Pro3.5.7 Year SetupThe Year Setup Screen is labeled “YEAr”.1. Enter the Setup Mode per Section 3.5.1. If necessary, rep

Page 15

Instruction 0024-939516Fyrite ProNotes:

Page 16

Instruction 0024-939517Fyrite Pro4.0 OPERATION4.1 Performing CO and Draft Testing4.1.1 Analyzer Turn On and Warm Up1. Connect the probe and make su

Page 17 - 3.4 Front Panel Push Buttons

Instruction 0024-939518Fyrite Pro4.1.2 Measuring CO (Model 100/105/120/125)1. Turn ON the analyzer and allow it to warm-up per Section 4.1.1.Importan

Page 18 - 3.5 Setup Mode

Instruction 0024-939519Fyrite Pro4.2 Performing a Combustion Test using the Model 110/120/1254.2.1 Analyzer Turn On and Warm UpImportant: The probe

Page 19 - 3.5.3 CO Channel Setup

Instruction 0024-939520Fyrite ProAtmospheric Burner or Gravity Vented Forced Air Hot Water Tank4.2.3 Sampling PointForced Air Furnace – When testing a

Page 20 - 3.5.6 Date Setup

Instruction 0024-939521Fyrite Pro90% Eff. Condensing Furnace/BoilerAtmospheric or Gravity Vented Boiler 80% Eff. Fan Assist or Power Vented Furnace/B

Page 21 - 3.5.7 Year Setup

Instruction 0024-939522Fyrite Pro4.2.4 Performing a Com bus tion Efficiency Test1. After turning ON the analyzer and selecting the appropriate fuel,

Page 22

Instruction 0024-939523Fyrite Pro4.3 Operating Tips• When an analyzer is brought in from a cold vehicle, let it warm up slowly to minimize condensatio

Page 23 - 4.0 OPERATION

Fyrite ProInstruction 0024-9395iContentsOrdering Information ... iii1.0 INTRODUCTION ...

Page 24

Instruction 0024-939524Fyrite Pro4.4 Using the BacklightThe LCD can be read in dimly-lit areas by pressing the LIGHT button. The backlight automatica

Page 25 - 4.2.2 Selecting a Fuel

Instruction 0024-939525Fyrite Pro2. Press the ▼ button once to display the Save Screen. The number shown in this screen represents the memory location

Page 26 - 4.2.3 Sampling Point

Instruction 0024-939526Fyrite Pro4.8 Printing Test DataTip: To avoid printing errors, it is important to select the correct protocol per Section 3.5

Page 27

Instruction 0024-939527Fyrite ProFigure 4-1. Printer Alignment & Sample Printout60°max.45 cm(18 in.)max.ENTERHOLDI/OLIGHT BACHARACH, INC. FY

Page 28

Instruction 0024-939528Fyrite Pro4.9 Clearing Saved Test Data (Model 105/110/120/125)Clear all saved test data as follows:Note: If no test data has b

Page 29 - 4.3 Operating Tips

Instruction 0024-939529Fyrite ProFigure 4-2A. Model 100 Run Mode Screen9COPPMCO4.11 Resetting the MicroprocessorIf the analyzer ‘locks up’ and canno

Page 30 - 4.5 Using the Probe

Instruction 0024-939530Fyrite ProFigure 4-2C. Model 110 Run Mode ScreensFuel Select:F1 = Natural GasF2 = Oil #2F3 = LPGF4 = KeroseneO2Ambient /Combus

Page 31

Instruction 0024-939531Fyrite ProFigure 4-2D. Model 120 Run Mode ScreensFuel Select:F1 = Natural GasF2 = Oil #2F3 = LPGF4 = KeroseneCO & O2CO Air

Page 32 - 4.8 Printing Test Data

Instruction 0024-939532Fyrite ProFigure 4-2E. Model 125 Run Mode ScreensFuel Select:F1 = Natural GasF2 = Oil #2F3 = LPGF4 = KeroseneCO & O2CO Air

Page 33 - (18 in.)

Instruction 0024-939533Fyrite Pro4.13 Screen Descriptions4.13.1 Warm-Up Screens As soon as the analyzer is turned ON, a se

Page 34

Fyrite ProInstruction 0024-9395ii4.7 Opening and Viewing Saved Test Data ... 254.8 Printing Test Data...

Page 35 - 4.12 Run Mode Screens

Instruction 0024-939534Fyrite Pro4.13.3 CO & Draft Screen (Model 105/125)The CO reading is a measurement of how much carbon monoxide in ppm is pr

Page 36

Instruction 0024-939535Fyrite Pro4.13.5 Fuel Screen (Model 110/120/125) The Fuel Screen is displayed after warm-up, and is where an operator chooses w

Page 37

Instruction 0024-939536Fyrite ProThe CO reading is a measurement of how much carbon monoxide in ppm is present in the flue-gas stream. Note, however,

Page 38

Instruction 0024-939537Fyrite Pro4.13.9 Temperature Differential Screen (Model 125)The Temperature Differential reading is the measured difference be

Page 39 - 4.13 Screen Descriptions

Instruction 0024-939538Fyrite Pro4.13.11 Ambient Air Temperature Screen (Model 110/120/125) The AMBIENT reading is the combustion-air temperature in

Page 40

Instruction 0024-939539Fyrite Pro4.13.13 Sensor Error Screen (Model 100/105/110/120/125) An O2 sensor error is displayed if the analy

Page 41

Instruction 0024-939540Fyrite ProNotes:

Page 42 - 4.13.8 CO

Instruction 0024-939541Fyrite Pro5.0 CALIBRATION & MAINTENANCEImportant: Fresh batteries should be installed, and the unit allowed to sta bi lize

Page 43

Instruction 0024-939542Fyrite Pro5.2 Ambient Temperature Calibration (Model 110/120/125)Material Required: Calibrated ThermometerProcedure:1. Enter

Page 44

Instruction 0024-939543Fyrite Pro3. Set the simulator to 32.0 °F (0 °C); then use the ▲ and ▼ buttons to set the dis played value to match the simula

Page 45

Fyrite ProInstruction 0024-9395iiiOrdering InformationFyrite Pro Confi gurationsModel 100 100 Printer Kit105 110 120 125 125 Printer Kit125 Printer/Oil

Page 46

Instruction 0024-939544Fyrite Pro6. Press ENTER to store these calibration values and move to the next cal i bra tion screen, or hold down ENTER for 2

Page 47 - Testing the Display Segments

Instruction 0024-939545Fyrite Pro9. Press ENTER to store the new calibration values and move to the next cal i bra tion screen, or hold down ENTER for

Page 48 - Calibration (Model 125)

Instruction 0024-939546Fyrite Pro6. Connect the tubing of the Calibration Kit to the GAS inlet of the an- a lyz er. Note that the gas fitting (Item 6

Page 49

Instruction 0024-939547Fyrite Pro5.7 Draft Calibration (Model 105/125)Material Required: • Bellows • Micromanometer Range: ±8 in. H2O column (±2

Page 50

Instruction 0024-939548Fyrite Pro5.8 Water Trap / Filter MaintenanceThe Water Trap / Filter Assembly removes water con den sate from the gas sam ple,

Page 51

Instruction 0024-939549Fyrite Pro5.9 O2 Sensor Replacement (Model 110/120/125)Be sure to perform all of the checks listed in Sec tion 4.13.13 to ensur

Page 52

Instruction 0024-939550Fyrite Pro5.10 CO Sensor Replacement (Model 100/105/120/125)Be sure to perform all of the checks listed in Sec tion 4.13.13 to

Page 53

Instruction 0024-939551Fyrite Pro5.11 Pump Assembly ReplacementMaterial Required: • Pump, P/N 0024-3048 • #1 Phillips ScrewdriverProcedure:1. Disa

Page 54

Instruction 0024-939552Fyrite ProFigure 5-4. Case Components & Plumbing SchematicCO SENSORGASKET(Model 125E only)OXYGEN SENSOR& SOCKETPRESSUR

Page 55

Instruction 0024-939553Fyrite ProFigure 5-5. PCB ComponentsCOSENSOR(Model 125E only)ToBATTERYToOXYGENSENSORToPUMPPRESSURESENSOR(RED WIRE)(RED WIRE)(RE

Page 56 - 5.10 CO Sensor Replacement

Fyrite ProInstruction 0024-9395ivNotes:

Page 57

Instruction 0024-939554Fyrite ProNotes:

Page 58

Instruction 0024-939555Fyrite Pro6.0 PARTS & SERVICE6.1 Replacement PartsDescription Part No.Carbon Monoxide Sensor ...

Page 59 - Figure 5-5. PCB Components

Instruction 0024-939556Fyrite Pro6.2.2 Optional AccessoriesAC Adapter (Battery Eliminator) ...0024-12

Page 60

Instruction 0024-939557Fyrite ProUnited States621 Hunt Valley CircleNew Kensington, PA 15068Phone: 724-334-5051Fax: 724-334-5723Email: help@MyBacharac

Page 61 - 6.2 Accessories

Headquarters:621 Hunt Valley Circle, New Kensington, PA 15068Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666Website: www.MyBacharach.co

Page 62 - 6.2.2 Optional Accessories

Instruction 0024-93951Fyrite Pro1.0 INTRODUCTION1.1 The Fyrite ProThe Fyrite Pro series of instruments are hand-held an a lyz ers that are designed f

Page 63 - 6.3 Service Centers

Instruction 0024-93952Fyrite Pro1.2 Operational OverviewPressing the I/O button turns the analyzer ON. Note that there is a warm-up period of either 1

Page 64 - Printed in U.S.A

Instruction 0024-93953Fyrite Pro2.0 TECHNICAL CHARACTERISTICSThe Analyzer Directly Measures and Displays:• Ambient Temperature .......