Bacharach Model 100 Specifications Page 52

  • Download
  • Add to my manuals
  • Print
  • Page
    / 64
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 51
Instruction 0024-9395
46
Fyrite Pro
6. Connect the tubing of the Calibration Kit to the GAS inlet of the an-
a lyz er. Note that the gas fitting (Item 6) is only required on Fyrite
Pro Models 105 and 125.
7. Adjust the regulator for ap prox i mate ly 2 SCFH of excess flow (see
Figure 5.1).
8. After the analyzer has stabilized (2 to 3 min utes), use the and
buttons to set the displayed value to match the CO con cen tra tion
stamped on the gas cylinder.
9. Press ENTER to store the new calibration values and move to the
next calibration screen, or hold down ENTER for 2 seconds to store
the new calibration values and enter the Run Mode, or press the
I/O button to exit the Calibration Mode and turn OFF the analyzer
without saving the changes.
10. Disconnect tubing from analyzer and turn off gas flow.
Figure 5-1. Calibration Kit Hookup
1
23 4 5
LEAVE TOP
PORT OPEN
Parts Shown:
1. Gas Cylinder
2. Regulator*
3. Tubing*
4. Tee*
5. Flowmeter*
6. Fitting, Gas*
* Contained in Calibration Kit
3
6
Page view 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 63 64

Comments to this Manuals

No comments