Maintenance
(replacing the filter and airflow indicator)
To protect the air pump from damage due
to foreign particles and moisture, replace
the filter as it becomes dirty. With mod-
erate use (15 to 30 minutes a day), it is
recommended that the filter be replaced
once a month. In dirty environments or
with heavy use, replace the filter more
frequently. Always replace the filter when
it is
visibly dirty or wet.
Replacing the Filter:
1. Remove the black rubber probe tip.
2. Pick out the filter with a pin or tweezers. A
fine screen will remain in the tip of the
probe.
3. Insert the new filter in the probe tip.
Make sure that the filter is firmly seated
against the fine screen.
4. Replace the rubber probe tip.
Replacing Airflow Ball Indicator:
1. Remove the clear plastic section of the
probe tip by gently pulling and twisting
the white grip.
2. Turn the probe tip upside-down and tap
on it to remove the old airflow indicator
ball.
3. Insert the new airflow indicator ball into
the tip.
4. Reattach the probe tip to the probe
assembly.
Filter
Rubber
Probe Tip
Airflow
Indicator Ball
Probe Tip
Page view 9