Setup Menu Function
Provides a list from which to select an inactivity (key
press) timeout for automatic shutdown. If no key
presses occur for the time specified, the Fyrite
InTech
initiates an automatic shutdown.
Use the UP () and DOWN () arrow buttons to scroll
through Inactivity Timeout options (
never [default],
20, 30, or 60 minutes
). Use the ENTER key to enable
the selected timeout.
IMPORTANT: The instrument overrides
the inactivity timeout
, cancels the
automatic shutdown
(that is, the
instrument remains ON), and restarts the
timeout countdown if:
any key is pressed,
CO is greater than 50 ppm, or
O
2
is less than 18.8 %.
Provides a list from which the user may chose a
minimum purge duration time (minimum length of
time that the pump continues to run) after shutdown is
initiated. Use a longer Post-Purge Period if the Fyrite
InTech
has been exposed to large amounts of CO gas.
Use ENTER to enable the selected Post-Purge Period.
“PURGING SENSORS” is displayed on the shutdown
screen if a Post-Purge Period is enabled.
Use the UP () and DOWN () arrow buttons to scroll
through Post-purge Period options.