If batteries are accidently installed in the wrong polarity, a positive
temperature coefficient (PTC) thermistor will protect the instrument’s
electronic circuity. The instrument will operate once batteries are
properly installed and the PTC thermistor is allowed to cool.
3.2 Probe Installation
Install the probe by sliding the end of its tubing over the gas inlet
port on the top right side of the instrument as shown in Figure 3. The
tubing may be difficult to slide over the gas inlet port of the unit for
the first time. This was done intentionally to allow for a snug fit. Use
a little dish washing liquid diluted in water or heat the end of the tube
in hot tap water to help it slide onto the port.
Figure 3. Probe Installation
3
GAS INLET PORT
LINE
DRYER
PROBE
HANDLE
When changing
probe tubes,
hand tighten nut,
and then use a
wrench to tighten
≈1flattoseal.