Page 1 - Combustion Gas Analyzer

Fyrite® ProCombustion Gas AnalyzerInstruction 24-9395Operation & Maintenance Rev. 9 – July 2004Product Leadership • Training • Service • Reliabili

Page 2 - WARRANTY

Instruction 24-93954Fyrite ProOperating Time ... Up to 18 hours continuous (pump running and backlight off)Warm Up Time

Page 3 - Contents

Instruction 24-93955Fyrite Pro3.0 SETTING UP THE ANALYZER3.1 Preliminary StepsBefore using the analyzer . . .• Check batteries (Section 3.2)• Connec

Page 4

Instruction 24-93956Fyrite Pro3.3 Connecting the Probe Hose and ThermocoupleThe analyzer’s model number determines which hose and thermocouple connec

Page 5 - Ordering Information

Instruction 24-93957Fyrite ProFigure 3-2A. Connecting the 19-7110 Probe (Model 100)GAS INLETSAMPLE-GASHOSEWATER TRAP /FILTER ASSEMBLYAC ADAPTERJACKPRO

Page 6 - Instruction 24-9395iv

Instruction 24-93958Fyrite ProFigure 3-2B. Connecting the 24-3004 Probe (Model 105)DIFFERENTIALPRESSURE HOSE(Optional)WATER TRAP / FILTERASSEMBLYPROBE

Page 7 - 1.1 The Fyrite Pro

Instruction 24-93959Fyrite ProFigure 3-2C. Connecting the 19-7111 Probe (Model 110/120)T-STACKGASINLETSAMPLE-GASHOSETHERMOCOUPLECONNECTOR(T-STACK)WATE

Page 8 - 1.2 Operational Overview

Instruction 24-939510Fyrite ProFigure 3-2D. Connecting the 24-3004 Probe (Model 125)T-STACKT-AIRROOM AIR /PRIMARY AIRTHERMOCOUPLECONNECTOR(T-AIR)DIFFE

Page 9

Instruction 24-939511Fyrite Pro3.4 Front Panel Push ButtonsNote that a push button may perform several functions, depending on the analyzer’s model n

Page 10 - Fyrite Pro

Instruction 24-939512Fyrite Pro3.5 Setup ModeThe analyzer is preset at the factory for the parameters shown below, but can be changed as described i

Page 11 - 3.1 Preliminary Steps

Instruction 24-939513Fyrite Pro3.5.3 CO Channel SetupThe CO Channel Setup Screen is labeled “CO.”1. Enter the Setup Mode per Section 3.5.1. If neces

Page 12 - Thermocouple

WARRANTYBacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufacture and will con

Page 13

Instruction 24-939514Fyrite Pro3.5.5 Time SetupThere are two Time Setup Screens, one for hours and the other for min-utes. Two bars appear above the

Page 14

Instruction 24-939515Fyrite Pro3.5.7 Year SetupThe Year Setup Screen is labeled “YEAr”.1. Enter the Setup Mode per Section 3.5.1. If necessary, repe

Page 15

Instruction 24-939516Fyrite ProNotes:

Page 16

Instruction 24-939517Fyrite Pro4.0 OPERATION4.1 Performing CO and Draft Testing4.1.1 Analyzer Turn On and Warm Up1. Connect the probe and make sur

Page 17 - 3.4 Front Panel Push Buttons

Instruction 24-939518Fyrite Pro4.1.2 Measuring CO (Model 100/105/120/125)1. Turn ON the analyzer and allow it to warm-up per Section 4.1.1.Important

Page 18 - 3.5 Setup Mode

Instruction 24-939519Fyrite Pro4.2 Performing a Combustion Test using the Model 110/120/1254.2.1 Analyzer Turn On and Warm UpImportant: The probe m

Page 19 - 3.5.3 CO Channel Setup

Instruction 24-939520Fyrite ProAtmospheric Burner or Gravity Vented Forced Air Hot Water Tank4.2.3 Sampling PointForced Air Furnace – When testing at

Page 20 - 3.5.6 Date Setup

Instruction 24-939521Fyrite Pro90% Eff. Condensing Furnace/BoilerAtmospheric or Gravity Vented Boiler 80% Eff. Fan Assist or Power Vented Furnace/Boi

Page 21 - 3.5.7 Year Setup

Instruction 24-939522Fyrite Pro4.2.4 Performing a Combustion Efficiency Test1. After turning ON the analyzer and selecting the appropriate fuel, pre

Page 22

Instruction 24-939523Fyrite Pro4.3 Operating Tips• When an analyzer is brought in from a cold vehicle, let it warm up slowly to minimize condensation

Page 23 - 4.0 OPERATION

Fyrite ProInstruction 24-9395iContentsOrdering Information ... iii1.0 INTRODUCTION ...

Page 24

Instruction 24-939524Fyrite Pro4.4 Using the BacklightThe LCD can be read in dimly-lit areas by pressing the LIGHT button. The backlight automaticall

Page 25 - 4.2.2 Selecting a Fuel

Instruction 24-939525Fyrite Pro2. Press the  button once to display the Save Screen. The number shown in this screen represents the memory location

Page 26 - 4.2.3 Sampling Point

Instruction 24-939526Fyrite Pro4.8 Printing Test DataTip: To avoid printing errors, it is important to select the correct protocol per Section 3.5.8

Page 27

Instruction 24-939527Fyrite ProFigure 4-1. Printer Alignment & Sample Printout60°max.45 cm(18 in.)max.ENTERHOLDI/OLIGHT BACHARACH, INC. FYRI

Page 28

Instruction 24-939528Fyrite Pro4.9 Clearing Saved Test Data (Model 105/110/120/125)Clear all saved test data as follows:Note: If no test data has bee

Page 29 - 4.3 Operating Tips

Instruction 24-939529Fyrite ProFigure 4-2A. Model 100 Run Mode Screen9COPPMCO4.11 Resetting the MicroprocessorIf the analyzer ‘locks up’ and cannot

Page 30 - 4.5 Using the Probe

Instruction 24-939530Fyrite ProFigure 4-2C. Model 110 Run Mode ScreensFuel Select:F1 = Natural GasF2 = Oil #2F3 = LPGF4 = KeroseneO2Ambient /Combusti

Page 31

Instruction 24-939531Fyrite ProFigure 4-2D. Model 120 Run Mode ScreensFuel Select:F1 = Natural GasF2 = Oil #2F3 = LPGF4 = KeroseneCO & O2CO Air F

Page 32 - 4.8 Printing Test Data

Instruction 24-939532Fyrite ProFigure 4-2E. Model 125 Run Mode ScreensFuel Select:F1 = Natural GasF2 = Oil #2F3 = LPGF4 = KeroseneCO & O2CO Air F

Page 33 - (18 in.)

Instruction 24-939533Fyrite Pro4.13 Screen Descriptions4.13.1 Warm-Up Screens As soon as the analyzer is turned ON, a serie

Page 34

Fyrite ProInstruction 24-9395ii4.7 Opening and Viewing Saved Test Data ...254.8 Printing Test Data...

Page 35 - 4.12 Run Mode Screens

Instruction 24-939534Fyrite Pro4.13.3 CO & Draft Screen (Model 105/125)The CO reading is a measurement of how much carbon monoxide in ppm is pres

Page 36

Instruction 24-939535Fyrite Pro4.13.5 Fuel Screen (Model 110/120/125)The Fuel Screen is displayed after warm-up, and is where an operator chooses whi

Page 37

Instruction 24-939536Fyrite Pro4.13.7 CO & O2 Screen (Model 120/125)The CO reading is a measurement of how much carbon monoxide in ppm is present

Page 38

Instruction 24-939537Fyrite Pro4.13.9 Temperature Differential Screen (Model 125)The Temperature Differential reading is the measured difference betw

Page 39 - 4.13 Screen Descriptions

Instruction 24-939538Fyrite Pro4.13.11 Ambient Air Temperature Screen (Model 110/120/125)The AMBIENT reading is the combustion-air temperature in ei

Page 40

Instruction 24-939539Fyrite Pro4.13.13 Sensor Error Screen (Model 100/105/110/120/125) An O2 sensor error is displayed if the analyzer d

Page 41 - Screen (Model 110)

Instruction 24-939540Fyrite ProNotes:

Page 42 - 4.13.8 CO

Instruction 24-939541Fyrite Pro5.0 CALIBRATION & MAINTENANCEImportant: Fresh batteries should be installed, and the unit allowed to stabilize at

Page 43

Instruction 24-939542Fyrite Pro5.2 Ambient Temperature Calibration (Model 110/120/125)Material Required: Calibrated ThermometerProcedure:1. Enter

Page 44

Instruction 24-939543Fyrite Pro3. Set the simulator to 32.0 °F (0 °C); then use the  and  buttons to set the displayed value to match the simulator

Page 45

Fyrite ProInstruction 24-9395iiiOrdering InformationModel100 100 Printer Kit105 110 120 125 125 Printer Kit125 Printer/Oil KitPart Number 24-8200 24-8

Page 46

Instruction 24-939544Fyrite Pro6. Press ENTER to store these calibration values and move to the next calibration screen, or hold down ENTER for 2 sec

Page 47 - Testing the Display Segments

Instruction 24-939545Fyrite Pro9. Press ENTER to store the new calibration values and move to the next calibration screen, or hold down ENTER for 2 s

Page 48 - Calibration (Model 125)

Instruction 24-939546Fyrite Pro6. Connect the tubing of the Calibration Kit to the GAS inlet of the an-alyzer. Note that the gas fitting (Item 6) is

Page 49

Instruction 24-939547Fyrite Pro5.7 Draft Calibration (Model 105/125)Material Required: • Bellows • Micromanometer Range: ±8 in. H2O column (±

Page 50

Instruction 24-939548Fyrite Pro5.8 Water Trap / Filter MaintenanceThe Water Trap / Filter Assembly removes water condensate from the gas sample, and

Page 51

Instruction 24-939549Fyrite Pro5.9 O2 Sensor Replacement (Model 110/120/125)Be sure to perform all of the checks listed in Section 4.13.13 to ensure

Page 52

Instruction 24-939550Fyrite Pro5.10 CO Sensor Replacement (Model 100/105/120/125)Be sure to perform all of the checks listed in Section 4.13.13 to e

Page 53

Instruction 24-939551Fyrite Pro5.11 Pump Assembly ReplacementMaterial Required: • Pump, P/N 24-3020 • #1 Phillips ScrewdriverProcedure:1. Disassem

Page 54

Instruction 24-939552Fyrite ProFigure 5-4. Case Components & Plumbing Schematic

Page 55

Instruction 24-939553Fyrite ProFigure 5-5. PCB ComponentsCOSENSOR(Models 100/105/120/125)ToBATTERYToOXYGENSENSORToPUMPT-STACK(Models 110/120/125)T-AIR

Page 56 - 5.10 CO Sensor Replacement

Fyrite ProInstruction 24-9395ivNotes:

Page 57

Instruction 24-939554Fyrite ProNotes:

Page 58

Instruction 24-939555Fyrite Pro6.0 PARTS & SERVICE6.1 Replacement PartsDescription Part No.Carbon Monoxide Sensor ...

Page 59 - Figure 5-5. PCB Components

Instruction 24-939556Fyrite Pro6.2.2 Optional AccessoriesAC Adapter (Battery Eliminator) ...24-1

Page 60

Instruction 24-939557Fyrite ProUnited StatesCalifornia7281 Garden Grove Blvd.,Suite HGarden Grove, CA 92841Phone: 714-895-0050Fax: 714-895-7950Email:

Page 61 - 6.2 Accessories

Headquarters:621 Hunt Valley Circle, New Kensington, PA 15068Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666Website: www.bacharach-inc.

Page 62 - 6.2.2 Optional Accessories

Instruction 24-93951Fyrite Pro1.0 INTRODUCTION1.1 The Fyrite ProThe Fyrite Pro series of instruments are hand-held analyzers that are designed for o

Page 63 - 6.3 Service Centers

Instruction 24-93952Fyrite Pro1.2 Operational OverviewPressing the I/O button turns the analyzer ON. Note that there is a warm-up period of either 10

Page 64 - Printed in U.S.A

Instruction 24-93953Fyrite Pro2.0 TECHNICAL CHARACTERISTICSThe Analyzer Directly Measures and Displays:• Ambient Temperature .......