Bacharach 2820 Specifications Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 28
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
CO
2
Analyzer 2810/2815/2820
Instruction 19-9222
10
3.7 Calibration
Calibration of a 2810/2815/2820 CO
2
Analyzer consists of rst performing
either an Air Calibration on a 2810/2815 analyzer that zeros the instru-
ment using fresh ambient air, or an N
2
Calibration on a 2820 analyzer
that zeros the instrument using 100% N
2
.
Factory calibration provides the best accuracy over the analyzer’s entire
detection range; however, the user can perform the optional Gas Cal-
ibration procedure that will improve the analyzer’s accuracy at a pre-
programmed calibration point of 2,500 ppm CO
2
for the 2810/2815, and
5% CO
2
for the 2820. Note that the pre-programmed calibration point can
be changed using BACH-COM software version 2.22 or higher.
It is recommended that an Air Calibration or N
2
Calibration be per-
formed at the beginning of each work day or each time the instrument
is switched OFF/ON. The analyzer should also be checked on a periodic
basis with a higher concentration of CO
2
to ensure that it still meets its
accuracy specication.
Follow the procedure under Section 3.7.1 to calibrate a 2810/2815 Low
Range (LR) Analyzer, or the procedure under Section 3.7.2 to calibrate a
2820 High Range (HR) Analyzer.
3.7.1 2810/2815 Calibration Procedure
3.7.1.1 Air Calibration
Perform this procedure in an uncontaminated enviroment such as out-
side fresh air, an empty room, or corridor where the air is assumed to
contain 340 ppm CO
2
.
Important! DO NOT perform an air calibration after sam-
pling high concentrations of CO
2
gas. Also, be careful that
your breath does not affect this procedure.
1. Switch ON the analyzer, and then allow it to sample fresh air for
5 minutes. During this time, the internal pump of a 2815 Analyzer
should be turned OFF.
CO2 LR ppm CO2
Analyzer 340
Page view 13